Ny

=

net counts of isotope

No

=

net counts of tracer isotope

P

=

Ay

=

activity of isotope per aliquot

Ay

=

activity per sample

V4

=

total sample volume

Vo

=

aliquot size used for the analyses

gmount of tracer isotope added, dpm

Gross Alpha and Beta (large area AC-23 probes and smail area SAC-4)

8) 5

Sop Ci funit

——eeeeo
U

Result:

Error Term:

+ 9

C

+

BeT

C

-—-

BeT

$$

(4-10)

1

(4-11)

rr
E
« 2.22

Result:

E

+ 2

Error Term:

+¢

2.22

=

(4-9)

{4-12}

pCi/swipe

Cc

(4-13)

7 +B

Radioactive Standard Sources. Radioactive standard sources were used to calibrate instrumentation

on a weekly basis.
counters.

An electroplated 90sr-90y standard was used for the calibration of beta

An electroplated 239py standard was used for calibration of alpha counters and the alpha

spectrometer. A mixed standard containing 238py, 239pu, 237Np was uSed for energy calibration of
the alpha spectrometer.

Parameters describing these sources are listed in Table 4-4,

Radioactive Standard Solutions. The radiochemical procedures utilized calibrated solution standards

as internal tracers to quantify the radionuclides of interest.

Other standard radionuclide solutions were used to make up spike samples for the quality assurance

program, as well as for calibration of the gamma and liquid scintillation counting systems.
Table 4-4 for specific parameters.)

128

(See

Select target paragraph3