eeeea

ce eee

Figure 7 shows the influence of exposing

depth-dose and biological effect.

a Masonite phantom to 2000 kVp x ray from @ single direction than when the

Tne next

exposure is bilateral with half of the dose given to each side (33).

yrtality of
had

exposure,

count im the

irth,

and

the

fe

Figure 4

Ue

is attenuated by

decreased dose as the beam is attenuated,
is very uniform deposition of

phantom.
of

In

it

the unilateral

inverse

square and

Thus

the bone marrow of large animals being exposed would have a progressively

there

owest white

uTe

falls off as

the case of

absorption so that the exit dose is about 45% of the entrance dose.

fifth <

ance of

the dose

In

the surface dose.

expressed as percent of

a

lealiie

e 3 plots

The

dose in the phantom was measured by Sievert fonization chambers and is

5 causes of
count

a gies:
it
Sahpe
aon ,
ee bo
ree ai

his

this

ne fraction of
Se
Subsequent :

great.

However, with bilateral exposure

energy

The biological consequences of

throughout

the

tissue equivalent

the different dose pattern are

It is of considerable importance to bear these differences in mind

when evaluating

therapy of radiation

injury and

trying

to. make

an anima}

experimentation as comparable as possible to an assumed real-life human
exposure,
Figure 8 shows a comparison of bilateral exposure to 4 Pi exposure
This situation

is

fallout irradiation with
exposure approaching 4

from @ planar source,
yas MOTE
amy cin with 8

Lin was mos t-4]
Se mortality

et al-

(32)

ys followed
cillin and .

Pi

important when

trying

its wide

in energy and

range

source.

Since

to evaluate

fallout

the

the

hazards of

radiation

radiativn is delivered

the usual narrow beam geonetry is not applicable,

such a diffuse 360 degree

field,

the

decrease

of duse

with deptin

is less pronounced than that resulting from a bilateral exp sure

in

In

tissue

to an

x-ray beam because fallout from inverse square is ain efrect neutralized.
For the same euergy,

the dose at the center of

the body is approximately

“50% higher thau would result from a given air dose with narrow beam
geometry.

Figure 8 further illustrates

the depth-dose curve from an

experimental situation using spherically oriented cobalt-60 sources with a
phantom placed at their center,

compared with a conventional bilateral

depth-dose curve obtained with a single Cobalt source (34).
tase,

In

the

tdeeer

the air dose is usually measured at the point subsequently occupied

the center of the proximal surface of
the source.

the patient or animal with respect

For the field case, all surfaces are "proximal" in

individual is the same.

It is

the

this air dose which is measured by field

struments; it does not bear the same relationships as

the surface dose

the depth dose as air dose measured in a "point source” beam in the
-10)1-

er

(33).

Select target paragraph3