4. Calibration
The first part of this section covers the preliminary circuit
adjustments used for setting-up # new unit or after major servicing,
and the second part describes the calibration procedure which should
be followed to check the instruments operation before a survey.
Figure 20 shows the location of the controls.

4.4.1

Preliminary Adjustments
1.

B supply - Adjust R-80 to obtain a B plus of 250 volts.

2.

Filter circuit adjustments,
a.

Introduce a 2400 ops signal to the input (pin 6 of J-1).

Capacitors C-6 and C~8 are adjusted to produce maximum
attenuation to the 24,00 cps signal as measured at the
grid of V-1A.

(Set the gain control (R-3) above minimum.

A convenient way of obtaining a 21,00 cps signal is to
connect pins 5 and 6 of J-1 together, and then pressing

the lock-on button. |
be.

Place a 7200 cps signal on the input and adjust R-10

and R-13 until maximum attenuation is reached (measure
at the grid of V-2A).

ce

Introduce a 60 ops signal and adjust R-15 and R-16 to
produce maximum attenuation as measured on the grid of

V=2A-.
3.

Connect the output of a Telepulse Receiving Station to the
Teslepulse Central Station input through a 100 to 1 attenuator.

k.

First Sehmtt trigger (V-3) sensitivity - Adjust the front
panel control (R-3) antil pulses of 10 volts amplitude appear

- 46 -