ware
a a
”
ree tt
»

>

aa

‘
Zactory with ene exception: The detent which
2ngaged the groove in the pocket chamber did
not guide the chamber in a positive manner
whenit was inserted in the tube. Unless an
operator has had considerable experience he
finds it extremely difficult to insert the chamber and turn it to its proper position.
Use of the dosimeters at Operation Green-

house indicated a great need for 2- and 5-r
instruments, which were not available. (It is
extremely difficult to accurately measure doses
for a 700-mr mission on a 10-r dosimeter, and
some accuracy is lost in reading doses for a
3-r mission.) Many of the missions required

exposures which exceeded the range of the 200mr dosimeters but were too low to read with
accuracy on a 10-r instrument.
2.9.2

Geiger Mueller Type Field Survey
Instruments

For TU 3.1.5 two different types of Geiger
Mueller (GM) field survey instruments were
supplied: 75 Victoreen 263B beta-gamma survey meters and 35 El-Tronics SGM-18A betagamma survey meters. GM instruments that

were serviced for other units included the
AN/PDR-8B.
The Victoreen 263B instruments gave reations noted below. Changes in manufacture as
a result of experience gained at Operation
Sandstone resulted in easy calibration and good
linearity of resultant calibration curves. The
parts are easily accessible for repair and
maintenance. The following suggestions for
improvementare offered: (a) Even though the

size of the case has been reduced from that
used with the 263A models, it still tends to be
top heavy and will fall over easily in a boat or
vehicle. The shape of the case could be improved, and the center of gravity should be
lowered. (b) The probe has been greatly improved over that of the 263A, but the beta
shield is still difficult to remove. This should
be given someattention. (c) The life of the
1B85 Geiger tubes was unsatisfactory, and
insufficient replacements were provided. (d)
The case is not waterproof, and considerable
moisture can accumulate inside the instrument,

a ah eee

position. To remedy this fault it was neces-

sary to replace C in several instruments. In
some cases the meter absorbed moisture, re-

quiring a drying-out process. In the case of
nine instruments the phone jack was wired incorrectly which resulted in no pulse in the ear-

phones, After 23 monthsof use the following

items represented replacements: thirty GM tubes,

fifteen 14-volt batteries, thirty-three 300-volt

batteries, eight 673-volt batteries, five amplifier
units, and two meters. Corrosion was evident on

someof the fittings, especially the metal portion of the carrying handles.
The-El-Tronics SGM-18A instruments had
several desirable design features; but most of

the instruments received required changes in

order to obtain good calibration, and some
trouble was experienced with all tubes. The
design of the case is good, and parts are easily
accessible for servicing. However, GM tubes
required considerable replacementas did
1U5 tubes. The instrument requires selected
NC51 neon tubes which are difficult to obtain
in quantities from normal stock supplies. The
compensating register Rg was not of high
enough valueto allow correction for normal

10,000-ohm potentiometers (Rg) were replaced

with 50,000-ohm units which permitted a wide

range of adjustment and resulted in correct
calibration. Apparently some attention should
be given to a redesign of circuits, especially
the functioning of the NC51 neontube.
While TU 3.1.5 was not equipped with
AN/PDR-8B GM instruments, considerable
servicing was done for other units equipped
with them. The halogen filled GM tubes show
great promise because of their long life and
apparent rugged construction. After the
modifications that are detailed below were
made, the operation was very satisfactory, and

the response was linear and could be directly

read for all ranges. However, trouble was ex-

—

perienced with ruptured tube windows when the
instruments were taken by plane to higher aititudes. Four instruments were found to have an
open circuit in the “crimp” wire lugs connecting the V201 GM tube cable to the terminal
board. This fault resulted in no response to

one exception: Moisture affects the 16 pf elecTr

trolytic condenser, C, which results in meter
readings even though the switch is in the off

component variations. For this reason all

sonably satisfactory service, with the excep-

resulting, in most cases, in excessive meter
readings. Such cases respond to drying with

N

radiation on the 0.5 and 5 mr/hr ranges. Twelve

instruments were found to have no response or

. ae

3

mr.

ee
he -

a

Select target paragraph3