WIRE

DISTANCE FROM
POWER TRANSFORMER

SIZES
Wire

TO LINE SWITCH

AWG

Inches

Diameter
Millimeters

For Line Voltage of 95 to 125:
Feet

Meters
0-15

#3

25

31-150

0-590

15.1-46

#0

35

9.9

151-200

46.1-61

#000

,40

10.5

Deviation

from

the

Recommended

Power

6,0

Supply: -

The objective of the recommended power supply is to deliver at the line terminals of the x-ray machine
sufficient power for proper operation. The maximum rated demand of the unit is 10 KVA at 90% power
factor, Any combination of generating equipment, power supply and feed lines which will deliver this
power with a voltage change between load and no load of not more than 9% or preferably 6%, may be
considered adequate,
IMPORTANT-- Effect

of

frequency

change

While this equipment is designed to operate at either 50-cycles or 60-cycles, the x-ray timer is driven
by a synchronous motor the speed of which is proportional to the frequency. The timer dial calibration
is correct only at 60-cycles. At 50-cycles the timer runs at 5/6 the speed of 60-cycle operation, hence

actual timing is 6/5 or 1.2 times that shown on the dial,

Proper recognition must be given this condi-

tion when setting up technics and utilizing x-ray tube ratings. Disregard of the correction factor may
result in over exposed films or an overloaded tube, See Operation Section for details,