EFFECTS OF EXPOSURE GEOMETRY

363

factors, in accordance with the theories of Spencer and Fano (/3, 15), as extended
and applied by others (74; also Loevinger et al. in reference 8).
Bomb, fallout y-radiation. The depth-dose curve obtained for exposure in a
y-ray field from fallout is shown as curve @ in Fig. 4C (the curve is truncated

and does not indicate the skin dose). This curve was measuredin a fallout field (1),
and the air dose is that measured by Sievert ionization chambers enclosed in
sufficient copper to exclude @-radiation. The phantom actually used was 25 ¢m in

diameter; the curve was flat throughout the phantom as measured with thin-

walled Sievert chambers that measured y-radiation, as well as energetic 3-radiation
if present. The flat central portion of the curve is, of course, due to y-radiation
only. The relatively high doses at the edges (as high on the surface as fifty times
the midline dose) resulted from addition of y-radiation and @- or very low energy
y-radiation, not measured by the y-ray survey instruments used to determine air
dose. It is apparent from the curve that the y-radiation dose throughout the
phantomis essentially constant, except at the skin surface.
It is possible to construet very approximately the depth-dose curve to be expected in the semi-infinite plane fallout situation, using a source spectrum for the
fallout field (1), approximate build-up factors (2) calculated from the theories of
Spencer and Fano (13, /6), and the geometrical considerations developed in the
present paper. The resulting curve is essentially flat as in the experimental curve;
however, the midline tissue dose with the ealeulated curve is approximately 75 %
of the entrance air dose. The explanation for this difference between the calculated
and observed curve is not apparent.
It is pointed out that with both initial and fallout y-ray exposures the dose is
essentially uniform as one goes from one end of the phantom to the other. This is
in contrast to all the laboratory geometries described and is approached only with
dq exposure.
DISCUSSION

Comparison of depth-dose patterns. In the preceding results, the marked differences

in tissue dose, obtained with different exposure geometries for the same air dose as
conventionally expressed, have been stressed. The large discrepancies possible
must be kept in mind when only the air dose is quoted oris available. It is seen
that no laboratory radiations as they have been employed quantitatively simulate
the initial or fallout y-radiations from the atomie bomb. Perhaps more striking
than the differences, however, is the marked similarity of the depth-dose patterns
for most of the exposure situations, and their essential identity if the artifact of
expressing dose in terms of that received by the air rather than the tissues could be
abandoned. The geometries fall into two basic categories—unilateral exposure, and
a second to include all the other types considered. With the exception of unilateral

S01 2492

Select target paragraph3