Appendix B
INSTRUMENTS AND DECAY
This appendix contains a detailed description of the detecting instruments of primary importance
to the project, together with certain early-time de ay curves used throughout the body of this

report.

This information is not currently available in the literature.

B.1 INSTRUMENTS
Since the GITR 103 is basic to the project, it is described in greater detail. This instrument
is a dose-increment recorder consisting of: (1) two concentric ionization chambers with recycling electrometers, (2) a magnetic-tape recorder, (3) a mechanical timer, and (4) a control
circuit and a battery power supply (Figures 1.4 and B.1). The externally mounted detector unit
is connected to the main instrument assembly within the case itself.

The detector consists of a low-range ionization chamber constructed around a high-range

fonization chamber, with each chamber connected to a recycling electrometer. These electrometer circuits use a CK 5886 tube connected as a cathode-coupled blocking oscillator with the
interelectrode capacity of the first grid below the predetermined triggering level of the positive

voltage shift on the first grid (Figure B.2). When predete-mined voltage level is reached, the
circult is triggered and generates a pulse of fixed amplitude at the cathode. The pulse causes

the first grid to conduct and to transfer a constant, predetermined charge to the chamber.
Simultaneously, the pulse is recorded on magnetic tape. The pulse terminates at the cathode
in approximately 500 usec, and the tube is left nonconducting with a negative voitage on the first
grid, thus completing the cycle.

The gamma dose increment required to discharge the ionization chamber its directly proportional to the amount of charge transferred to the chamber (Figures B.3 and B.4). The charge
transferred during each cycle is constant but dependent upon the triggering level of the electrometer, which is controlled by the adjustable bias voltage of the second grid. Calibration of detectors is achieved by adjustment of the bias voltage until a predetermined dose increment causes
the electrometer to cycle (Section C.1). The calibration control for each chamber is located
on the moistureproof electrometer housing attached to the base of the chamber assembly.
The ionization chambers are constructed of thin-walled spun-aluminum shells mounted concentrically. Cylindrical and hemispherical surfaces are used wherever possible to establish
optimum voltage gradients for efficient charge collection. The chambers are filled with pure

argon at 7.5 psi and sealed by softsoldering over nickel-plated surfaces. The volumes of the
two chambers are 1,475 and 14.0 cm’ for the low-range and high-range chambers, respectively.
The sensitivity ratio of 1,000 between the two ranges is achieved by the design value of the input

capacity of the electrometer circuits. A lead-tin filter over the entire outer surface of the
detector provides uniform energy response from about 100 kev to 2 Mev (Figure C.1).
The record is made on 900-foot lengths of instrumentation-quality magnetic tape spooled
on standard 5-inch reels.

The tape is 0.25 inch wide and has a polyester backing 0.001 inch

thick. A Brush Electronic Company BK 1303-1 three-channel recording head, driven to tape

saturation, records unidirectional pulses on the tape. The maximum usable pulse packing is
400 pulses per inch of tape. Recording intervals of 12 hours and 60 hours_are used with tape
transport speeds of 0.25 in/sec and 0,05 in/sec, respectively. These speeds are accurate to
+ 2 percent for the entire recording interval. Both recorders are of identical construction
with the exception of the drive motors. A single 6.7-volt mercury-battery stack having a

359

“A nk
cw
OF

Select target paragraph3