eo il CRER-1068 APPENDIX 1 Determination of Total Strontium and Barium Dry sample at 100°C, then ash at 600°C for 24 hours in a muffle furnace. Fs Weigh 100-200 mg of ash into a polythene capsule. b) Insert sealed silica tubes containing strontium and barium standards into the sample (preparation of standards described in Appendix 2 Irradiate sample and standards for two hours in a slow neutron flux of about 2 x 1012 neutrons/cm@/sec. Dissolve the irradiated ash in a Pyrex beaker containing 30 mg strontium chloride, 10 ml of 7 M hydrochloric acid and one drop of nitric acid. to promote exchange. Heat to boiling for ten minutes. Pass solution through a Dowex-1l anion exchange column containing 2 ml of resin (as received). of 7 M hydrochloric acid. Wash with 5 ml Evaporate column effluent to dryness. Dissolve solids in 2 ml of water. Add 8 ml of .yellow fuming nitric acid, 1.49 ~ 1.50. specific gravity pee Reaieeamnrete it + * Cool in cold water bath for three minutes. barium nitrates precipitate. Strontium and 10. Centrifuge and discard supernatant liquid containing calcium. li. Repeat steps 7, 8, 9 and 10 twice. le. Dissolve precipitate in 6 ml of-water. 13. Add 4 drops of iron carrier solution (2 mg Fe-+/ml). 14. Heat in boiling water bath, then precipitate ferric hydroxide by bubbling in ammonia gas. Centrifuge. 15. Add 2 more drops of iron carrier, and centrifuge precipitate on top of first. 16. Decant supernatant liquid into a clean centrifuge tube.