de 1. O-4 mr/hr 2. 0-40 mr/hr 3. 0-400 mr/nr 4. 0-4000 mr/hr 5. 0-40,000 mr/hr i Bs The microphonios to be held to a minimum. On any one of three one-inch drops produced by suddenly pulling a one-inch board from under one end of the instrument the meter reading would not go over one-half full scale on any range of sensitivity. e. The instrument was to be calibrated by gamma rays from radium and to conform to the sensitivities specified in (c) above within an average of 10% of full scale at any point on the scale. f. The circuit to be so designed that after switching from one range to another not more than 10 seconds would be required to reach 90% of the final reading. ge The instrument to be sensitive to gamma radiation and the wall | -of the chamber to be composed of a material with the atomic number less than ten such that the instrument would be wave length independent from all X and gamm radiation with energies above 5 Kvp, and beta particles with energies less than 1.0 NEV to be excluded. he The circuit time constant to be such that 90% of the final reading of the instrument would be reached within 10 seconds or less. 4. The instrument to operate under the above requirements over ranges of temperatures from minus 5° C to 70° C and with the relative humidity of 95% for a period of twenty-four hours. ,Seotion IX 18 ee,