gamma-emitting 85g. tracer is determined using a NaI (Tl) crystal and muicichannel analyzer. DETERMINATION OF PLUTONIUM IN URINE, WATER AND MISCELLANEOUS SAMPLES 3Y ALKALINE-2ARTH PHOSPHATE PRECIPITATION Reagents Sodium Nitrate Oceyl Alconol Nicric Acid Phosphoric Acid Potassium Hydroxide Hydrochloric Acid Cone. and 7.2N 85% GN Conc. Calcium Nitrate Saturated Solution (Filtered) Eluting Solution 30 ml HC1, 0.3 mi HF/Liter H,0 Hydrogen reroxide Anion Exchange Resin Sodium 31sulfate Ammonium Hydroxide 30% AGLx4 50-100 mesh Sis 58% Sodium Sulfate 15% (Filtered) l 4 d/n/ml oe 242pluronium Tracer Plutonium is co-precipitated with urine salts by alkaline earth phosphates. The organic material carried by the precipitate is dry ashed in a muffle furnace. Plutonium and urine salts are dissolved in 7.2N nitric acid. The pluto- mium fraction is absorbed onto an anion exchange resin and eluted with 0.36N HCl - 0.008N HF. Plutonium is electrodeposited onto %"" diameter stainless steel discs and its activity determined by alpha pulse height spectrometry. The srocedure is as follows: 1. add sample to an appropriate size beaker recording aliquot volume. Rinse sample container with 7.2N HNO, 2. 3 and add to sample beaker, add an additional 5 ml of cone. HNO,, place sample on a stirring hot plate and adjust temperature to 80° + 5°C, 12