SEPARATION OF 7°STRONTIUM FROM URINE SAMPLES 1 LITER OR LESS Reagents Strontium Tracer Octyl Alcohol Nitric acid Strontium Carrier Cone. 20 mg/ml Calcium Chloride Oxalic Acid Sodium Hydroxide Hydrochloric Acid Aydrochloric Acid 01M Saturated Solution 6M Conc. 0.08N Yttrium Carrier 4 (2 . s Tne procedure 20 mg/ml is as follows: measure sample into a 1.5 liter Seaker, place beaker on a stirring hot place and heat slowly to 80-85°C, acidify sample to pH 1 with nitric acid (add acid in small amounts to prevent excessive foaming, use a few drovs of octyl alcohol if neces- sary), add 40 mg each Sr carrier and Y carrier, 1 ml 8° 5. tracer and 50 ml O.1M Cacl., digest with stirring at 80-85°C for 30 minutes, adjust to pH 4 with 6M NaOH, add 40 ml saturated oxalic acid soiution and mix well, teadjust to pH 4 with 6M NaOH and digest, with stirring, at 80-85° for 30 minutes, remove from heat, remove stirring 2ar and let settle overnight, 10. filter sample through a Whatman #42 ashless filter paper using dilute NH,OF wash solution to rinse beaker and precipitate,