184

FOOTE

for data accumulation. Each counting period was followed by data read-

out on punched paper tape and initiation of the subsequent counting

period. All data were tape-to-card converted and supplied to a computer for réduction. The field-instrumentation enclosure with its detector pod is shown in Fig. 1. An example of the change in time of the

Fig. 1—-Field instrumentation for total-absorption gamma-ray spectrometer.

field gamma-ray energy spectral shape caused from decay of surface

nuclear fallout is given in Fig, 2.
Proper and adequate methods of data reduction mustthen beutilized to determine what radionuclides provide the gammaradiation of

which the field spectrum is composed. The precise energy spectral
calibration standard of each radionuclide composing the field gamma‘ray spectrum must be determined. The optimum computer method for

the determination of the amounts of each radionuclide must be resolved

on the basis of the computer time requirement and results. Energy

calibration of all spectral data is required. The radionuclides that
create the most intense gamma-ray energy lines can initially be lo-

cated. The selection of two energy lines such as the 2.615-Mev line

from 787] and a line of lesser energy, preferably of low energy, allows

energy calibration to a high degree of accuracy and consistency.! There
are many ways of locating the photopeak energy positions, Texas In-

struments’ method of locating is indicated in Fig. 3, where the sides of
a near-Gaussian-shaped curve are matched by two straight lines; the
16h