EFFECTS OF EXPOSURE GEOMETRY

61

Rotational exposure.® This type of exposure in which the sourceis fixed and the
phantom is allowed to rotate on its vertical axis can be regarded as the limiting

case of multilateral exposure, and the curve essentially superimposes on that for

multilateral exposure (curve d, Fig. 44). Thus, there is no advantage of this
type of exposure over the bilateral or multiport geometries. Identical results are
obtained if the phantom is held constant and the source is allowed to revolve
about the phantom at a constant TSD of 100 em.

Crossfire technique. With the crossfire technique, only a single exposure from two

opposing “point” sources energized simultancously is used, as opposed to the
bilateral technique in which two exposures, first one side and then the other, are
made with a single source. The resulting dose pattern is shown as curve a, Fig.
“4B. It is apparent that the shape of the curve is negligibly different from that
obtained with bilateral, multilateral, or rotational techniques, and that the tissue
dose is still considerably below the air exposure dose that the phantomis said to
have received.
The reason for the lowtissue dose relative to air dose may not be immediately
apparent, since with crossfire technique the air dose threughout the exposure
volume is essentially constant. It is easily seen, however, if one considers that as
soon as the animal or phantom is introduced the ‘skin’ dose at either side (and

throughout the phantom) inumediately drops considerably because of absorption
in the tissue or phantom. Thus, the entire curve is well below the entrance air dose.
The crossfire curve is higher than the bilateral curve because of what might be
regarded as an artifact of dosimetry resulting from the manner in which azr dose
is measured with the two techniques. This can be seen as follows: With the bilateral
technique, the total air “dose”? administered is the sum of two entrance air doses
from the two half-exposures. With the crossfire technique, the total air “dose”
given is the sum of the entrance air dose from one machine and the exif air dose
from the opposite machine (less by inverse square). Thus the total air ‘‘dose’” with
crossfire measured at either surface of the exposure volume (1 or B, Fig. 1), is less
than with bilateral, and the tissue dose, in terms of the per cent of air ‘‘dose,” is
correspondingly greater. It should be noted that exposure with crossfire for one
half the fofal time for both Lalf-exposures with bilateral (two tubes on simultaneously with crossfire) yields a tissue-dose curve that superimposes on the bilateral
curve. Since, as noted, the air dose for the same total time is less with crossfire,
however, the exposure time with crossfire for the same total air ‘dose’? is longer
‘This method of exposure should be clearly differentiated from the multiple-port or rotational exposure used in radiotherapy of tumors. In the elinic, a collimated beamis emploved
which at any given time exposes, in theory, only the tumor mass and a small volume of over-

Iving skin and tissue at any instant. Thus, with multiple-port or rotational therapy, the deep
tumor, always in the field, receives a maximum dose and any given portion of overlying skin
receives a minimum dose.

Select target paragraph3