lo

EXPERIMENTAL SIMULATION AND CEOMETRY FACTOR
In such a diffuse field, the decrease of dose with depth in tis-

sue is less pronounced than that resulting from a bilateral exposure
to an Xeray beam and the relationship to airedose differs as noted in
the two cases, The result is that, for a given energy, the dose at

the center of the abdomen is considerably higher than a given proximal

air-dose would imply for the narrow-beam or point-source case,

Figure 7,1 illustrates the depth dose curve in a 36-cn diameter
cylindrical masonite phantom from an experimental simulation of the

fieldpee (reference 13) using a spherically oriented group of

36 Co™

sources,

The phantom was placed at the center of the assembly.

This is compared to a conventional bilateral depth-dose curve measured

in the same phantcm and obtained with a single Co

source,

Both are

normalized to air~dose, but the average air-dose at all points later
occupied by the phantom surface is implicit for the diffuse case, while
the proximal air-dose is used in the Lilateral case,

Figure 7.2 is a similar comparison for 200-KVP, 0.S=mm, copper-=

filtered Xerays, with the diffuse geometry that of a plane rather than

spherical source assembly.

This wes produced in this case by rotation

of the phantom and ion chamber in the beam of a stationary Xeray unit.

The useful beam angle of the unit was wide encwgh to include the whole

phantom,

The averape air-dose around the circumference was here used

for the diffuse geometry and the proximal airedose again in the bilateral exposure,

It is evident that for both these energies (the effec-

tive energy of the X-ray beam being atout 90 KV), the diffuse-narrow
beam depth dose ratio for either 2 m radians (plane) or hw steradians
(volume) diffuse geometry is almost the same, That is, the midline
dose is about 50 percent higher and the Seem dese is 35 percent higher

than the same airedose (measured proximally) woulc imply in the narrow
beam bilateral exposure, It is therefore assumed that this approximate

factor will apply throughout the field exposures,

On this tasis the air-dose values calculated from the survey meter

the situation to that of a bilateral exposure to a source with the same
energy distribution but using a point source geometry and a proximally
measured air-cose, Alternatively, if a point source of higher energy,

say C060, were used bilaterally in the same way to simulate a field

exposure to only the higher gamma components, then the meter energy
correction factor woulc be unity, In this case, to specify a bilateral
exposure yielding a midline dose equal to that with diffuse geometry,
the point source air-dose should be the diffuse field air-cose meas-

ured with the meter anc multiplied by (1.09 x 1.5) only.
The doses are discussed further in Chapter 8,

s2.

ei9

age eee apereerenee

readings (Table 8.1) should be multiplied by 1,5 in order to compare

Select target paragraph3