2.4.13

UW-1
The UW-1 underwater instruments functioned excellently.

When check calibrated, they were found to retain accuracy with no
drift.

2,4,14

Rad-Chem Trailer Equipment
The radiation measuring equipment in the Rad-Chem trailers

functioned adequately.

The Berkley Model 2000 scalers presented

some trouble, but were generally accurate and reliable.

These units

were calibrated and tested with a Rutherford Model 300 pulse generator using a 0.25 volt,

1ms rise time,

2 msduration, negative pulse,

The best GM tube was the Tracerlab TGC-2 with a window
thickness of less than 2 mg/cm’, which gave counting efficiencies of
from Z0 to 30 percent,
The alpha counter head and amplifier-power supply system
functioned well for low level alpha counting, providing good discrim-

ination against beta and gamma pulses.
The laboratory monitor, SU-3B, was used to monitor bench
tops and miscellaneous items.
Initially the voltage regulation at the Site Elmer trailer was
poor, due to a defective Sorenson voltage regulator.

The voltage reg-

ulation stability was carefully examined with a fast trigger oscilloscope,
and line disturbances were damped out by careful tuning to resonance,